Basic performance and reference of the connector salt spray machine [Figure]

If the salt spray test chamber shows the wrong symptoms and the staff knows nothing about it, let it continue to work, the chances of destroying the machine will increase. So what are the symptoms of the salt spray test chamber failure? Can you take those methods to solve the problem? Let's take a look at the common troubleshooting tips.

A. Over-temperature indicator: There are over-temperature indicators on both the left and right sides (the test room on the left and the saturated tank on the right). When the overheating light is on, there are three cases.

1. Safety temperature controller alarm device setting error

2. If the over-temperature indicator is continuously lit, notify the company to process

3. Check if the temperature setting is too low, please reset

B. Low water level indicator: When this light is on, the operation power will be cut off. At this time, water should be added automatically in the test room or saturated barrel until the low water level indicator goes out (check if the water source of the water inlet is open).

C. End indicator: When this light is on, it means the test time ends.

Judgment and treatment of abnormal function of salt spray test chamber

Condition 1: The test chamber cannot rise to the set temperature

Cause: The temperature of the laboratory temperature controller is set too low;

2 The laboratory safety switch setting is too low.

Treatment method: 1. Set the temperature controller to the required temperature;

2. Set the safety switch to the desired temperature.

Condition 2: Insufficient amount of spray

Reasons: 1. The spray regulator is placed too low;

2. The glass filter s in the preheating tank is blocked;

3. The pressure setting is too low.

Treatment: 1. Increase the spray regulator;

2. Clean the glass filter:

3. Adjust the pressure regulating valve to a pressure of 1 kg/cm2, and the air compressor is marked with a pressure regulating valve adjusted to a pressure of 2 kg/cm2.

Condition 3: When the spray volume is too high

Reasons: 1. The spray regulator is placed too high;

2. The glass nozzle is used for a long time, and the nozzle diameter is too large.

Treatment: 1. Use a spray regulator to lower;

2. Replace the new glass nozzle.